Dental handpiece maintenance

FanEliza 发布

Dental handpiece maintenance note tips:

1 When using the handpiece, it must be lubricated strictly according to the instruction manual. Generally, the high-speed handpiece is lubricated twice a day. Low-speed handpiece (straight machine, curved machine, motor) once a day.

2 The working pressure of the dental handpiece must be strictly in accordance with the instructions. In general, the working pressure of two-hole and three-hole dental handpiece is 2-2.2Kg, and the working pressure of four-hole, five-hole, and six-hole dental handpiece is 2.4---2.5Kg (refers to the pressure at the interface between the handpiece and the connecting pipe). The pressure must be adjusted when the new treatment machine is used, and the pressure of the treatment machine must be re-calibrated after a period of use (about one month).

3 Disinfection: High-temperature steam disinfection should be used, chemical liquids and dry heat disinfection are not allowed. The disinfection temperature is 135℃, and the maximum time is no more than 30 minutes. It must be cleaned before disinfection and lubricated after disinfection. Note: The dental handpiece with quick-change connector should be removed. Only the dental handpiece is disinfected, and the connector is not disinfected. For low-speed dental handpiece, remove the straight and curved heads for disinfection. When there is dirt on the surface of the dental handpiece, please clean it with alcohol.

4 Do not use standard, rusty or aged burs or grinding heads.
5 Press the cover high-speed dental handpiece must not press the cover when rotating, so as not to damage the movement. Straight and curved machines cannot be loaded or removed when they are rotating.
6 Burs, grinding heads or "standard rods" should be taken and installed once a day to avoid rusting if the burs are not taken out for a long time.